8 TECHNICAL SPECIFICATION
8.1 Standard Specification
8.2 Common specification
DBC 150-TC and TS DBC 650-TC and TS
Temperature range -45°C (from ambient) to +150°C 50°C to 650°C
Well depth dry well 160mm (6.3in)
insertion depth 155mm (6.1in)
dry well 160mm (6.3in)
insertion depth 155mm (6.1in)
Well diameter 30mm (1.18in) 30mm (1.18in)
Stability 0.03°C 0.05°C (between 70°and 650°C)
0.15°C (between 50°and 70°C)
Radial Uniformity
(adjacent holes)
0.15°C 0.25°C
Heating time 14 minutes (20°C to 120°C) 22 minutes (25°C to 650°C)
Cooling time 22 minutes (25°C to -20°C) 27 minutes (600°C to 100°C with
optional cooling probe)
Display resolution 0.01°C 0.01°C
CJ compensation error 0.3°C 0.3°C
Reference probe accuracy 0.2°C (including B1probe) 0.3°C (including B2 probe)
Power supply 100-120 or 220-240 Vac,
50-60 Hz, 230 VA
100-120 or 220-240 Vac,
50-60 Hz, 1385 VA
DBC 150 TC only DBC 650 TC only
Maximum mA IN (current signal) 35 mA 35 mA
Maximum V IN (voltage signal) 30 Vdc 30 Vdc
Display area 60 x 40mm (2.36 x 1.57 in) LCD with backlighting
Temperature units °C
K
°F
External Ref. Probe Facility for connection of a PT100 reference probe
Switch Test Facility for connection of switch contacts
Communication Interface RS232, supplied as standard
Software function Ramp, step and preset temperature generation. Switch test
Operating temperature range 10 to 50°C
Reference Standards
PT 100
{DBC-TC only} Thermocouples
EN 60751: 1998-05 (ITS 90)
EN 60584-1: 1997-10
EN 60584-2: 1998-02
Conformity Standards
EMC
Safety
EN 61326 (1997) + A1(1998) + A2 (2001)
EN 61010-1: 2001 (2nd edition)
Measurement Category I (all measurement terminals MAX 30V to earth)
Pollution degree 2
Regulation sensor Thermocouple
Weight 9.5 kg (21 lbs)
Dimensions 322 x 156 x 328mm (12.7 x 6 x 12.9 in)