3.4 Temperature Reference Selection
The DBC series measures the dry well temperature in two different ways:
Internal reference probe
From the standard working screen, press the menu key to open the menu selection
screen:
Using the keys, move the selection cursor on the REF. T display. Select
INTERNAL with the keys and press ent to activate the dry well temperature
measure made with the internal regulation probe. The next display allows the
setting-up of the temperature generation mode (Direct, Ramp, Step or Preset).
External reference probe
From the REF. T display of the Menu Selection Screen select the SEL RTD function
using the keys and press ent.
The next selection display now appears as follows.
MEASURE
INTERNAL
SEL RTD
NONE
RTD
mA
T/C
REF T.
MEASURE
INTERNAL
SEL RTD
NONE
T/C
RTD
mA
REF T.