6 FAULT FINDING
Note:
Fuses are designed to give maximum protection in case of fault. Make sure
that voltage selector and fuses are selected for the nominal working value.
Fault description Possible problem Solution
1. The instrument does not
work even if the power
cord is correctly fitted and
the mains switch is in the
correct position.
- faulty protection fuse
- faulty power cable
- faulty power switch
-Replace the fuse
-Replace the power cable
-Contact your service representative
2. When the instrument is
turned on the fuse blows
- short circuit on power supply card
- faulty power switch
-Contact your service representative
3. The heating and cooling
function of the dry well is
not working
- faulty power supply card
- faulty controller card
- damage to one or more Peltier cells
-Contact your service representative
4. The controller is not
disabled when the set-
point is reached
- faulty controller card
-Contact your service representative
5. The heating and cooling
times out of specification.
-damage to one or more Peltier cells or
damage to the heating system
-Contact your service representative
6. Fan system not working. -faulty power supply card
-faulty controller card
-faulty fan
-Contact your service representative
7 At high temperatures the
instrument cannot
stabilize.
-incorrect step or preset values
-faulty reference probe
-Change the step or preset values
-Contact your service representative