The display shows the mode settings (i.e. how the RTD measurement can be
displayed).
•Direct
This screen shows, in the top, a direct reading in the selected temperature unit,
together with the RTD type and the temperature. The bottom part of the screen
shows, the generated temperature value (with internal or external reference), the
temperature set-point, the status arrows symbols and the selected generation
mode.
• CAL Mode
This facility performs an error calculation analysis for the unit under test. From the
MODE display select CAL using the and press ent. The next display provides
the setting of RTD span.
Using the select the T min value and set the temperature limit using the
numeric key-pad: then, move the cursor on the T max value using the key, set
the related temperature value and press ent to confirm. Press ent again to show
the next display, where the error calculation method can be set.
MODE
DIRECT
CAL
Pt-100
RDG: °C
0,15% SPAN
148.50
150.00 DIRECT
BLK
°C
148.00
SET SPAN
T min : 50.00 °C
T max : 200.00 °C