Nominal Value is the recommended value for the test): pressing the ent key causes
the microprocessor to combine the current indications on the display (True Value
and Effective Reading).
Calibration of the resistance (Ohm) input ranges (DBC-TC only)
As for the millivolt input range, also the resistance measurement channel has
different ranges capability, depending on the RTD measurement available.
From the CALIBRATION menu select the required Ohms input range to be
calibrated, using the keys, and press ent.
Using a certified and high accuracy Ohms simulator, apply the required value to the
related electrical connection on the front panel of the instrument (the nominal value
is the recommended value for the test).
Pressing the ent key causes the microprocessor to combine the resistance
measure on the display (true value and effective reading).
CALIBRATION
Ohms INPUT
1 KOhms INPUT
Ohms INPUT
Nominal value:
300
True value: 300.00
Reading: 298.00
Press ENT or CE ...