Agilent Technologies 10737R Washer/Dryer User Manual


 
Chapter 7O Agilent 10737L and Agilent 10737R Compact Three-Axis
Interferometers
Procedure
7O-20 Users Manual
Figure 7O-9. Agilent 10737L Compact Three-Axis Interferometer
return beam pattern
12 Install the receiver assembly.
To do this, reverse the Removing the receiver assembly procedure,
above.
13 Plug in the fiber-optic cables.
14 Adjust each receivers gain by turning its gain adjustment screw to
cause the receivers LED to light, then reduce the gain until the LED
just turns off. For more information, see Agilent 10780F instructions in
Chapter 8, Receivers, of this manual.