A SERVICE OF

logo

4.2.3 Special parameters
The following parameters are identified as special parameters.
Unless particularly instructed by the Mitutoyo service personnel, do not modify the default
settings. Unnecessary modification may result in a counting error.
Parameter
No.
Parameter Function Setup Data Initial Value
16 Counting range setting 0: Normal
1: Expand
0 (Normal)
17
Display mode at start-up selection
Selects the display when the power is turned on
(when the main power switch on the rear panel
is turned on, or when the power to the counter
is turned on after power failure).
Select whether to display or
numbers. (If 0 is set for Parameter 15, 0.000 is
displayed, and if 1 is set for Parameter 15, the
saved value is displayed.)
0:
display
1: Number
display
1 (Number
display)
60 Self-check 0:
1:
2:
0
92 Decimal-point position
(Select the number of digits below decimal point.)
0: 0 digit
1: 1 digit
2: 2 digits
3: 3 digits
4: 4 digits
5: 5 digits
6: 6 digits
2 (2 digits)
93 Number of divisions 0: 4
1: 8
2: 20
3: 40
4: 80
5: 100
6: 125
7: 200
8: 250
9: 400
2 (20)
94 Detection factor 0: 1X
1: 2X
2: 5X
3: 10X
0 (1X)
No. 99MBE032A