2. SETUP
2.3 Connecting to the Counter
When connecting the power cord and external equipment to this counter, connect the
connectors to the rear panel as shown below. (Below diagram is applied for 2-axis
counter.)
IMPORTANT
When connecting to external equipment, take note of the following items.
• Make sure the main power switch on the rear panel is off.
(This is not the display ON/OFF switch on the front panel.)
• Use the supplied power cord only.
• Connect this counter with the machine tool using the supplied grounding wire to avoid a
risk of electric shock.
• The power-supply line for this counter should be separate from that for the machine tool,
and if possible should be supplied from a dedicated outlet.
• Make sure the cables and external equipment are connected to the counter securely.
• If the Linear Scale is not connected to this counter, the least significant digit may flicker,
but this is not a malfunction.
• Since the scale model (AT100 series or AT700 series) is identified at a timing when the
main switch is turned to ON, always turn the main switch to ON with the desirable scale
being connected.
No. 99MBE032A