TRIGER send a trigger to the Power Module.
Syntax TRIGER <ch>
Parameters the number of the Power Module to send the trigger
to.
Category Trigger and Sync Type: Overlapped
VALL? return load voltage measurements of all installed Power Modules. The
response is a series of values - one for each installed Power Module,
separated with “,”, starting with the lower numbered module. If auto-
polarity is enabled, negative output values will be preceded with a
minus sigh.
Syntax VALL?
Parameters none
Category Read-back Type: Sequential
VHIGH program the upper window warning level for the load voltage of the
Power Module.
Syntax VHIGH <ch>,<value>
Parameters the number of the Power Module to be programmed.
the value to be set as the upper window warning
level for the load voltage. The upper limit of the
programmed value is VMAX. The lower limit of the
programmed value is VSET.
Category Warnings Type: Sequential
VHIGH? return the current settings for the upper window warning level of the
load voltage of the Power Module. The response is a <value>.
Syntax VHIGH? <ch>
Parameters the number of the channel to be queried.
Category warnings Type: Sequential
VLIM program an upper limit on subsequent programming of voltage
settings for the Power Module.
Syntax VLIM <ch>,<value>
Parameters the number of the Power Module to be programmed.
the value to become the new upper limit for VSET
programming. The upper value of the programmed
value is VMAX.
Category Settings Type: Sequential
XMP 2600 Programming Manual
rev. 1.1 65