This chapter introduces some of the advanced features of the XMP 2600 and provides
guidelines on the usage of those features.
Workpoint window warning
Power Modules of the XMP 2600 can monitor their outputs and warn the user if the
output values exceed a defined range. The window warning thresholds that define the
aforementioned range have both a high value (Vhigh for voltage and Ihigh for current)
and a low value (Vlow for voltage and Ilow for current).
Usage
guidelines
The following sections describe how to use the workpoint window warning feature of the
XMP 2600.
Setting the window thresholds
Programming the window thresholds is done using the following commands:
VHIGH <ch>,<value> for the high voltage
threshold
Vmax
Imax
Imin
Vhigh
range
Vlow
Ilow
Ihigh
range
Iset
V
IHIGH <ch>,<value> for the high current
threshold
Vset
VLOW <ch>,<value> for the low voltage
threshold
I
ILOW <ch>,<value> for the low current threshold
The high threshold values can range from the set point (Vset or Iset) to the rating value
(Vmax or Imax).
The low threshold values can range from 0 to the set point value (Vset or Iset).
Enabling the window thresholds
The high and low thresholds can be enabled or disabled, in pairs, using the commands:
WHIGH <ch>,{1|0} (use 1 to enable and 0 to disable)
and
WLOW <ch>,{1|0} (use 1 to enable and 0 to disable)