3. DISPLAYS AND KEY OPERATIONS
3.2.14 Group judgment
While the tolerance judgment is applied to each measurement from a workpiece, this group
judgment is applied to a group of the specified number of workpieces.
Judgment to Group 1 Judgment to Group 2 Judgment to Group 3
Individual
judgment
1
Individual
judgment
2
Individual
judgment
3
Individual
judgment
4
Individual
judgment
5
Individual
judgment
6
Individual
judgment
7
Individual
judgment
8
In the basic setup select whether group judgment is to be performed. If it is, then set the
group size (the number of workpieces included in a group), calculation items (mean,
maximum value, minimum value, and range), and group lower limit and upper limit. If
“Not performing group judgment” is selected, the setup guidance for it will not be dis-
played.
The group judgment will be in effect in a single-run measurement or continuous-run
measurement.
a) For the result display and GO/NG judgment indication each individual measurement
and judgment result will be used.
b) RS-232C output
In the basic setup it is possible to set whether the group judgment result data is output-
ted for the RS-232C interface. If it is, the output contents from the group judgment will
be as follows:
P0, ( GO) 12.34567 ... Individual data
P0, ( GO) 12.34560 ... Individual data
P0, (+NG) 12.34600 ... Individual data
GP0, ( GO) 12.34575 ... Group judgment result data
Each individual piece of measurement data can be the objective of statistical processing,
however, group measurement data will be excluded from statistical processing.
Even if “Err-0” (specified workpiece not present) occurs, the obtained data will not be
cleared. To abort the measurement, press the
C
key (or input the “CL” command
via the RS-232C interface).