3 - 8
No. 99MBC032A
5) Sample measurement
A measurement where the number of samples is set to “2~999” is called a “sample
measurement”.
In practice this will take place as a single-run measurement or a continuous-run mea-
surement.
From the measured data the calculation items (mean, maximum value, minimum value,
and range) that have been set for the sample measurement will be calculated and the
resulting data will be automatically subject to GO/NG judgment. In addition, the
measured data will be outputted for the RS-232C interface, Digimatic Output Unit
interface, and printer.
The measured data are available for statistical processing.
This is suitable for run-out measurement and cylindricity measurement.
6) Statistical processing
Measured data from single-run and continuous-run measurements can be statistically
processed (i.e. the number of measurement times, standard deviation, maximum value,
minimum value, mean, and range are calculated).
These statistical processing results can be outputted for the display, printer (statistical
memory for all programs will be cleared after printout), and RS-232C interface.
Press the
STAT
/
S.E
key (or input “ST” command via the RS-232C interface) to start
statistical processing, and press it again (or input the “NST” command via the RS-232C
interface) to terminate statistical processing.
Press the
A.CL
/
M.CL
key to clear the statistical memory of the foreground program
(case of a simultaneous measurement), and press the
SHIFT
and
A.CL
/
M.CL
keys to
clear the statistical memory of all the programs.
These statistical results data will be stored in memory while the power is on, and will
be lost when the power is turned off.