![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/f/18/f18ddd3c-603d-4b2c-8f84-e3b488149268/f18ddd3c-603d-4b2c-8f84-e3b488149268-bg74.png)
6 - 6
No. 99MBC032A
6.1.3.2 Setting the LSM-9506
The following setting for LSM-9506 needs to be set in the Basic Setup mode and Setting Up
the Functions mode.
1) Basic Setup (Refer to the section 4.1, “Basic Setup” and 4.1.2.5, “Selecting and setting the
function in the B4 mode”.)
a. Setting the use of the RS232C port RS-232C : PRN
b. Setting the RS232C communication baud rate BAUD : 9600
c. Setting the RS232C communication data bits LENGTH : 8
d. Setting the RS232C communication parity bit PARITY : NONE
e. Setting the delimiter for communication DELIMT : CR+LF
f. Setting the RS232C line control CONTRL : USE
2) Setting Up the Functions (Refer to the section 4.5, “Setting Up the Functions” and 4.5.3.6,
“F5: Setting the data output conditions”.)
DAT O.C : 1 (Printing for each measurement)
DAT TIM : 0 (Printing all)
Connect with the printer after setup is complete. Press [Print] key to start measurement and
printing measured data.
6.1.4 RS-232C commands
• This section describes the reception commands used by the RS-232C interface and the
response (transmission) commands that correspond to the reception commands. The data
section of each command consists of ASCII codes.
• Use the following command subscriptions, symbols and supplement descriptions when you
read Section 6.1.5, “List of commands”.
• Command symbols
Symbol
Meaning
Entry of numeral
Combination using numbers between 1 and 7
Select a number between 1 and 255
However, aaa should not be identical to bbb.
Segment number
Start edge and finish edge of the edge specification
aaa: Start edge, bbb: Finish edge
Setup data or measured data
Measurement interval number
Data output condition
Periodical output timer value (sec)
Invalidation period of automatic workpiece detection (msec)
Number of measurement times of sample measurement
Number of measurement times of automatic workpiece detection
Number of scans for averaging
Group size subject to judgment
Space character
Maximum of 7 significant digits
Select a number between 1 and 4
Select a number between 0 and 9
Select a number between 0 and 999
Select a number between 0 and 9999
Select a number between 0 and 999
Select a number between 1 and 999
2 (select between n=0 and 11)
Select a number between 0, 1, 2 and 99
sssssss
aaa-bbb
±ddd.dddd
Program Number Select a number between 0 and 9p
m
r
nnn
ttt
nn
SHL (Threshold level) specification (%)
Select between 50% and 90%pp
mmm
nnnn
Δ
n
tttt