A SERVICE OF

logo

AZI Customer Service 800-528-7411 or 602-470-1414 Page 87 of 92
13. MAX 5000XL SPECIFICATIONS
Feature
Specification
Size
18.5” length x 12.7”width x 9.5” height
470 mm length x 323 mm width x 240 mm height
(Height varies slightly with adjustable feet for leveling purposes.)
Weight
31 lbs. (14 kg)
Casework
Textured aluminum castings
Power Requirements
100-130 VAC, 50/60 Hz @ 8 Amps
220-240 VAC, 50/60 Hz @ 4 Amps
Fused on/off switch as part of an EMI power entry module
Operating
Environment
0-35º C @ 50% humidity
0-30º C @ 80% humidity
Non-condensing, non-explosive
Moisture Range
0.1 99.9%
Heating Range
25º C to 600º C
Heater Calibration
Menu Driven
NIST Traceable using an accessory Temperature
Calibration Interface module
Calibration tolerance:
± 1.5º C up to 275º C
± 2º C up to 600º C
Twelve calibration points from 50º C to 575º C.
Twelve preset verification points from 50º C to 575º C, or
up to four user-programmable verification points.
Programmable calibration reminder
Temperature Control
± 1º C at temperatures between 50º C and 300º C
± 2º
C at temperatures between 301º C and 600º C
Test Temperatures
User programmable through front panel keypad to control
idle temperature, test temperature, and high start
temperature for all tests.
Multiple test temperatures are programmable for linked
test sequences
Heater Chamber
Composite ceramic fiber lined heater chamber
Isolated from electronics to minimize damage from
corrosive sample materials
Oven self cleaning cycle (550º C for 45 minutes)
Balance
Integrated balance with automatic operation
Balance capacity to 100 grams
Resistant to nominal vibration, with options available for
isolation feet
Resistant to air currents from internal exhaust fan
Isolated from extreme temperatures in the test chamber
Isolated from extreme changes in humidity caused by
evaporation in the test chamber