Microplate Luminometer Maintenance
103
b) Connection at injector head
Measurement head Injector head
Turn MONOLIGHT 3096 off and disconnect it from the wall
socket.
Open instrument door of MONOLIGHT 3096, open screws of
service door and take door off (Figure 8-9).
Push tubing fitting screw over the tubing such that the
threading is facing the tubing end (Figure 8-10). Then push a
washer over the tubing.
Using the special tool supplied, expand the tubing end by
pushing the tubing over the tool’s arbor.
Pump T-piece Washer Banjo bolt Tubing
Push T-piece into expanded tubing end.
Insert tubing end with T-piece into threaded borehole of pump
head, turn fitting screw with washer into the threading of the
pump head and finger-tighten it.
Figure 8-9:
MONOLIGHT 3096
open
Figure 8-10:
Manufacturing the
tubing connection