Chapter 7O Agilent 10737L and Agilent 10737R Compact Three-Axis
Interferometers
Procedure
7O-20 User’s Manual
Figure 7O-9. Agilent 10737L Compact Three-Axis Interferometer —
return beam pattern
12 Install the receiver assembly.
To do this, reverse the “Removing the receiver assembly” procedure,
above.
13 Plug in the fiber-optic cables.
14 Adjust each receiver’s gain by turning its gain adjustment screw to
cause the receiver’s LED to light, then reduce the gain until the LED
just turns off. For more information, see Agilent 10780F instructions in
Chapter 8, “Receivers,” of this manual.