A SERVICE OF

logo

Pressure Systems, Inc. Model 9116 User’s Manual
Page 84 www.PressureSystems.com
ii dd Description
‘q’ read
index
0C
00
01
Set Cal Valves to RUN or LEAK Position (default)
— choice made by ii=12.
Set Cal Valves to CAL/RE-ZERO or PURGE Position
choice made by ii=12.
0B
see chart
below
0D-0E -- Reserved for factory use
0F
00
01
Disable periodic Thermal Coefficient Update task.
Enable periodic Thermal Coefficient Update task (default).
10
01-20 Set Number of A/D Samples to Average, (default = 8).
Valid values are 4, 8,16, 32, and 64. Other values below 64
are rounded up to the next valid value listed above.
05
11 -- Reserved for factory use
12
00
01
Set Cal Valves to RUN or CAL/RE-ZERO Position (default)
— choice made by ii=0C.
Set Cal Valves to PURGE or LEAK Position — choice made
by ii=0C.
0B
see chart
below
13
00
01
Use Static IP Address Resolution (default)
Use Dynamic IP Address Resolution (RARP/BOOTP)
(Results in immediately becoming the module’s new power-on
default)
06
14
00
01
02
Disable Host Response/Stream Back-Off Delay (default).
Enable Host Response/Stream Back-Off Delay as low-order
byte of Ethernet Address(converted to decimal value * 20
sec.).
Enable Host Response/Stream Back-Off Delay specified
per eeee as decimal value (* 20 sec.)
07
16
00
01
Disable Host Response/Stream Total Size Prefix (default).
Enable Host Response/Stream Total Size Prefix (2-byte
big-endian binary value with total size of response or stream
data in bytes that follows it).
08
17
00 Set TCP Connect Port per eeee as decimal value
(default=9000).
09
18
00
01
Disable Auto UDP Broadcast at Reset (default).
Enable Auto UDP Broadcast at Reset.
0A
19
00
01
Set Minimum Temperature Alarm Set Point (in
N
C)
per eeee as decimal value (default = 0
N
C).
Set Maximum Temperature Alarm Set Point ( in
N
C)
per eeee as decimal value (default = 60
N
C).
0D
0E
1B
00 Set Thermal Update Scan Interval per eeee as decimal
value (seconds), 1 <= eeee <= 3600 seconds (default = 15).
11