Pressure Systems, Inc. Model 9116 User’s Manual
Page 30 www.PressureSystems.com
RESET (Command ‘B’)
Purpose: Instructs the module to reset internal operating parameters, and to set all internal
control variables to their default “reset” state (see description below). The current
TCP/IP socket connection will remain open. Execution after a power off/on cycle
is optional (unnecessary).
Command
“B”
‘B’ is the command letter.
Response
“A”
‘A’ is the acknowledge letter.
Description: The module returns to the following “reset” states if this command is executed:
● Re-zero correction (offset) terms are set to the last values stored in
transducer memory.
● Span correction (gain) terms are set to the last values stored in transducer
memory.
● Calibration Valve is set to the RUN Position
● Number of Samples for Data Averaging is set to last value stored in non-
volatile memory (factory default = 8).
● Any autonomous host data delivery streams defined by ‘c’ sub-commands
are reset (undefined).
● The Multi-Point Calibration function defined by ‘C’ sub-commands is reset
(undefined) if in progress.
Example:
● Send TCP/IP command to a Model 9116 module (via open socket) to reset defaults:
“B”
Read following response:
“A”