Agilent E1441A SCPI Command Reference 75
Chapter 3
Comments • You can substitute decimal values for the OFF (“0”) and ON (“1”) parameters.
• The function generator calibration is secured when shipped from the factory
and the security code is set to "HP_E1441".
• *RST does not change the state.
Example Set the calibration state to unsecured:
CAL:SEC:STAT OFF,HP_E1441
Unsecure function generator calibration.
:SECure:STATe?
CALibration:SECure:STATe? returns a "1" or "0" to show whether the calibration
security state is enabled (1) or disabled (0). The number is sent to the output buffer.
Example Query the calibration security state:
CAL:SEC:STAT?
Query function generator calibration
security state
enter statement
Enter value into computer.
:SETup
CALibration:SETup <0 | 1 | 2 | 3 | . . . | 84> configure the function generator's
internal state for each of the calibration steps to be performed.
:SETup?
CALibration:SETup? queries the calibration setup number. Returns a value
between 0 and 84.
:STRing
CALibration:STRing <quoted string> allows you to record calibration
information about your function generator while CAL:SEC:STAT is OFF. For
example, you can store information such as the last calibration date and/or the next
calibration due date. The calibration message may contain up to 40 characters. The
string is stored in non-volatile memory.
Parameters
Comments • The calibration message can contain up to 40 characters.
• Calibration security state must be OFF to store a string.
Parameter Name Parameter Type Range of Values Default Units
quoted string discrete alphanumeric none