A SERVICE OF

logo

98
IMPORTANT
For better accuracy it is recommended to calibrate the instrument frequently.
Calibration should be performed whenever the pH electrode is replaced, after
performing a cleaning procedure or changing the reference electrolyte, and at
least once a month.
INITIAL PREPARATION
Pour small quantities of pH 7.01 (HI 7007) and pH 3.00 (HI 5003) solutions
into clean beakers.
For accurate calibration, use two beakers for
each buffer solution; the first for rinsing the
electrode, the second for calibration.
Note: The electrode should be submerged at least 4 cm (1½") into the solution.
PROCEDURE
Note: If you need to calibrate to NBS standards, use pH 6.86 (HI 7006)
solution instead of pH 7.01.
Use the supplied HI 98501 thermometer as reference.
Switch the instrument on by pressing the ON/OFF
button.
Rinse and immerse the pH electrode in pH 7.01 buffer
and shake briefly. Wait a couple of minutes for the
reading to stabilize.
Note the temperature of the buffer solution using the
supplied HI 98501 thermometer (e.g. 20°C).
Press the °C key to select temperature setting, then adjust the Temperature
knob until the LCD shows the noted temperature.
pH CALIBRATION
Press the pH key to select pH measurement.
Wait a couple of minutes, then adjust the OFFSET knob until the display
shows the pH value at the noted temperature (see the "pH versus
temperature" chart at page 10).
Rinse and immerse the pH electrode in pH 3.00 buffer and shake briefly.
Wait a couple of minutes, then adjust the SLOPE knob until the display
shows the pH value at the noted temperature (see the "pH versus
temperature" chart at page 10).
The pH calibration is now complete.
4 cm
(1½")
HI 7007
RINSE
HI 7007
CALIBRATION