National Instruments SCC-LP02 Dryer Accessories User Manual


 
National Instruments™, NI™, ni.com™, and NI-DAQ™ are trademarks of National Instruments
Corporation. Product and company names mentioned herein are trademarks or trade names of their
respective companies. For patents covering National Instruments products, refer to the appropriate
location: Help»Patents in your software, the patents.txt file on your CD, or ni.com/patents.
© 2002–2004 National Instruments Corp. All rights reserved.
371071B-01 Mar04
*371071B-01*
Table 1 lists the signal connection corresponding to each pin. AI (X) and
AI (X+8) are the analog input signal channels of the E Series DAQ device.
AI GND is the analog input ground signal and is the reference for AI (X)
and AI (X+8). A GND is the reference for the ±15 V supplies. AI GND and
A GND connect to the SC-2345 at the SCC-PWR connector. You can use
pins 17 to 20 for cascading channels. Refer to the Device Specific
Information section for more information on cascading configurations.
Table 1. SCC-LPXX Pin Signal Connections
Pin Number Signal
1 E Series AI (X)
2 E Series AI GND
3
4 E Series AI (X+8)
5
6 E Series AI GND
7
8 E Series AI GND
9
10
11 A GND
12
13 +15 V
14 –15 V
15
16
17 AI (X)– (from first stage)
18 AI (X+8)+ (from first stage)
19 AI (X)+ (from first stage)
20 AI (X+8)– (from first stage)